• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 1708160
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during preparation of a port placement procedure, wrong device was allegedly sent in the package.There was no patient contact.
 
Additional Manufacturer Narrative
As this malfunction is considered one event; only one mdr report will be submitted for the reported quantity affected of 23 for this event.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation as well as photos were provided for review.The investigation of the reported event is currently underway.(expiration date: 04/2023).
 
Event or Problem Description
It was reported that during preparation of a port placement procedure, the wrong type of port was allegedly sent.There was no patient contact.
 
Additional Manufacturer Narrative
H10: as this malfunction is considered one event; only one mdr report will be submitted for the reported quantity affected of 23 for this event.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: twenty two powerport isp implantable port kits were returned for evaluation.Gross visual evaluation was performed.However, two electronic photos were provided for review.The investigation is confirmed for the reported component misassembled issue, as each powerport isp implantable ports within the kits were noted to have bumps throughout the port septum.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 04/2023), h11: g1, h6(result, conclusion), h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 8F
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14139907
Report Number3006260740-2022-01358
Device Sequence Number14488781
Product Code LJT
UDI-Device Identifier00801741026676
UDI-Public(01)00801741026676
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K072215
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1708160
Device Catalogue Number1708160
Device Lot NumberREFY2010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/24/2022
Supplement Date Received by Manufacturer05/26/2022
Initial Report FDA Received Date04/19/2022
Supplement Report FDA Received Date05/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-