• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. VENTRIO MESH; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. VENTRIO MESH; SURGICAL MESH Back to Search Results
Catalog Number 0010211
Medical Device Problem Codes Defective Device (2588); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Abscess (1690); Adhesion(s) (1695); Bacterial Infection (1735); Cellulitis (1768); Pain (1994); Hernia (2240); Insufficient Information (4580)
Date of Event 07/02/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges "past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient"; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Should additional information be provided, a supplemental emdr will be submitted.Not returned.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol ventrio on (b)(6) 2014.As reported, the patient is making a claim for an adverse patient outcome against the ventrio.Attorney alleges general allegations for "past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient." it is also alleged that the patient experienced emotional distress and the device was defective.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol ventrio on (b)(6) 2014.As reported, the patient is making a claim for an adverse patient outcome against the ventrio.Attorney alleges general allegations for "past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient." it is also alleged that the patient experienced emotional distress and the device was defective.Addendum per additional information provided: (b)(6) 2014 - patient was diagnosed with ventral hernia in left lower abdominal wall thereby underwent open repair with the implant of ventrio mesh.Per operative notes, ¿the hernia sac was encountered and then opened.The small hernia above and medial was closed with sutures and a piece of ventrio mesh was placed into the defect with a smooth side down and a rough side up.¿ (b)(6) 2015 - patient was diagnosed with cellulitis and abdominal wall abscess thereby underwent incision and drainage of left lower quadrant abdominal wall abscess.(b)(6) 2015- patient was diagnosed with wound infection, infected mesh and recurrent ventral hernia thereby underwent repair with mesh removal.Per operative notes, ¿an abundant area of granulation and chronic abscess cavity was encountered.The mesh contained granulation and some purulent material.Adherent omentum with adhesions were noted.The ventrio mesh was completely excised and a piece of synthetic mesh was placed and sutured.¿ attorney alleges that the patient had abscess, adhesions, pain, infection, hernia recurrence and emotional injuries.It is also alleged that the patient had stress, anxiety, permanent physical limitations in regard to exercise and physical scaring.
 
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges "past, present, and future damages, including but not limited to, mental and physical pain and suffering for severe and permanent personal injuries sustained by the patient"; however, no details have been provided.No lot number has been provided; therefore, a review of the manufacturing records is not possible.Addendum: this supplemental emdr is submitted to document additional information provided.Based on the additional information received, there is no change to initial determination, no conclusions can be made.Per medical records review, about 9 months post implant of ventrio mesh, patient was diagnosed with abscess, hernia recurrence, cellulitis, granuloma, bacterial infection and adhesions thereby underwent repair with mesh removal.The instructions-for-use supplied with the device list adhesions, hernia recurrence and infection as possible complications.The warning section of the ifu states, "if an infection develops, treat the infection aggressively.Consideration should be given regarding the need to remove the mesh.An unresolved infection may require removal of the device." review of the manufacturing records was performed and found that the lot was manufactured to specification.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRIO MESH
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC.
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED 3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
4018258495
MDR Report Key14139996
Report Number1213643-2022-93218
Device Sequence Number6311436
Product Code FTL
UDI-Device Identifier00801741016387
UDI-Public(01)00801741016387
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K100229
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2015
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/28/2015
Device Catalogue Number0010211
Device Lot NumberHUWD1224
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/12/2022
Supplement Date Received by Manufacturer03/08/2024
Initial Report FDA Received Date04/19/2022
Supplement Report FDA Received Date03/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/27/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Disability;
Patient Age37 YR
Patient SexFemale
Patient Weight105 KG
-
-