Model Number N/A
Medical Device Problem Codes Break (1069); Fracture (1260); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported, that: during primary hip replacement taperloc complete with allofit cup, the ceramic biolox head broke by hitting the head on the stem.Correct technique and instruments are used.Patient involvement - no further outcome.
 
Additional Manufacturer Narrative
(b)(4).Initial report.Report source, foreign - event occurred in the (b)(6).The client has indicated that the product is in process of being returned to zimmer biomet for an investigation.The investigation is currently underway.Once the investigation has concluded, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d4, g3, g6, h1, h2, h3, h6, h10 visual examination of the returned product confirmed that the head has fractured into seven pieces.The are several metal transfer marks visible on the fractured pieces.This device is used for treatment.The likely condition of the device when it left zimmer biomet is conforming to specification.The root cause of the reported event cannot be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTA CER FEM HD 32/+3MM T1
Common Device Name
BIOLOX DELTA CERAMIC FEM HD
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14140350
Report Number3002806535-2022-00222
Device Sequence Number20391300
Product Code LZO
UDI-Device Identifier00887868271557
UDI-Public(01)00887868271557(17)311214(10)3095624
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K200959
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number650-1161
Device Lot Number3095624
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2022
Supplement Date Received by Manufacturer07/01/2022
Initial Report FDA Received Date04/19/2022
Supplement Report FDA Received Date07/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-