This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4, b5, g1, g2, g3, g6, h1, h2, h3, h6, h10 *check udi, & d10* visual examination of the returned product identified signs of use and the tip was fractured.The zrm suggested that the device might experience a bending overload fracture.Review of the device history record identified no deviations or anomalies during manufacturing.Medical records were not provided.The complaint is confirmed.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after the investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|