MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number MI1000 MED-EL CONCERT PIN
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Conclusions: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.This is a combined initial and final report.
 
Event or Problem Description
Initial information about a possible re-implantation has been received.The explanted device has been received for investigation on (b)(6), 2022.As of (b)(6), 2022 no further information has been received, despite request.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key14141143
Report Number9710014-2022-00275
Device Sequence Number17202910
Product Code MCM
UDI-Device Identifier09008737062750
UDI-Public(01)09008737062750
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/10/2022
Device Model NumberMI1000 MED-EL CONCERT PIN
Device Catalogue Number07680
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2022
Initial Report FDA Received Date04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/10/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age5 YR
Patient SexFemale
-
-