MAUDE Adverse Event Report: THOMMEN MEDICAL AG CONTACT MC INICELL, PF 4.5, E 3.5, L12.5; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THOMMEN MEDICAL AG CONTACT MC INICELL, PF 4.5, E 3.5, L12.5; DENTAL IMPLANT Back to Search Results
Model Number 4.23.635
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 09/14/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Primary stability not achieved, implant was completely covered with bone, no thread tap used, bleeding.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTACT MC INICELL, PF 4.5, E 3.5, L12.5
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
THOMMEN MEDICAL AG
neckarsulmstrasse 28
grenchen 2540
CH  2540
Manufacturer (Section G)
THOMMEN MEDICAL AG
neckarsulmstrasse 28
grenchen 2540
CH   2540
Manufacturer Contact
astrid weber
neckarsulmstrasse 28
grenchen 2540
CH   2540
MDR Report Key14141159
Report Number3003184527-2022-00849
Device Sequence Number3986052
Product Code DZE
UDI-Device Identifier07640156470550
UDI-Public07640156470550
Combination Product (Y/N)N
Initial Reporter CountryOM
PMA/510(K) Number
K090154
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/09/2025
Device Model Number4.23.635
Device Lot Number22357
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2022
Initial Report FDA Received Date04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age58 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-