| Model Number |
309635 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Address information was not able to be obtained, therefore, (b)(6) was used as a place holder.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd syringe with hypodermic needle precisionglide¿ there was a mix of product.There was no report of patient impact.The following information was provided by the initial reporter: it was reported that only syringe in the lot with a red cap.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary: one sealed packaged syringe and one photo of a 5ml luer-lock syringe (p/n 309635) batch 1274106 with red needle attached was received and evaluated.The packaged sample has an 18x1 ½ inch blunt fill needle attached to the syringe in a 20x1 ½ inch package.The condition observed is non-conforming per product specification.Potential root cause for the incorrect needle defect is associated with the assembly process.As a corrective actions, quality deficiency notice was given to a responsible associate for line clearance and quality alert will be issued to the entire plant for awareness.A device history record review showed no rejected inspections or quality issues during the production of the provided batch number that could have contributed to the reported defect.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd syringe with hypodermic needle precisionglide¿ there was a mix of product.There was no report of patient impact.The following information was provided by the initial reporter: it was reported that only syringe in the lot with a red cap.
|
| |
|
Search Alerts/Recalls
|