Model Number AR40E
Medical Device Problem Codes Break (1069); Inaccurate Delivery (2339)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis therefore, a visual analysis of the complaint device cannot be completed.A review of the device/lot history record and historical data analysis for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch report will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported when surgeon was advancing the intraocular lens (iol) he noticed the trailing haptic had a kink, lens was partially inserted but then retracted.Patient contact was confirmed.Through follow-up, additional information was received stating no unplanned incision enlargement, no serious patient injury, no unplanned suture(s), and no unplanned vitrectomy.The same procedure was completed using back-up lens with the same model and diopter size.Patient outcome post-procedure was reported as great.No further information is available.
 
Additional Manufacturer Narrative
Additional information was received with device return: section d-11 concomitant product: emeraldc cartridge lot: cj35992.Product investigation was completed, and the following fields were updated accordingly: section d-9: device available for evaluation? yes.Section d-9: returned to manufacturer: yes.Section d-9: date returned to manufacturer: 06-may-2022.Section h-3: device evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed that the lens was received with one haptic bent and twisted.The lens was cleaned, and creases could be found on the optic body, consistent with a lens that was folded.No issues could be identified that would contribute to a delivery issue.The complaint issue of haptic damaged was identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14141593
Report Number3012236936-2022-00886
Device Sequence Number20391309
Product Code HQL
UDI-Device Identifier05050474502000
UDI-Public(01)05050474502000(17)260909
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/09/2026
Device Model NumberAR40E
Device Catalogue NumberAR40E00150
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer05/25/2022
Initial Report FDA Received Date04/19/2022
Supplement Report FDA Received Date06/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EMERALDC CARTRIDGE LOT: CJ35992
Patient SexFemale
Patient Weight82 KG
Patient RaceWhite
-
-