| Model Number |
AR40E |
| Medical Device Problem Codes |
Break (1069); Inaccurate Delivery (2339)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis therefore, a visual analysis of the complaint device cannot be completed.A review of the device/lot history record and historical data analysis for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch report will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported when surgeon was advancing the intraocular lens (iol) he noticed the trailing haptic had a kink, lens was partially inserted but then retracted.Patient contact was confirmed.Through follow-up, additional information was received stating no unplanned incision enlargement, no serious patient injury, no unplanned suture(s), and no unplanned vitrectomy.The same procedure was completed using back-up lens with the same model and diopter size.Patient outcome post-procedure was reported as great.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received with device return: section d-11 concomitant product: emeraldc cartridge lot: cj35992.Product investigation was completed, and the following fields were updated accordingly: section d-9: device available for evaluation? yes.Section d-9: returned to manufacturer: yes.Section d-9: date returned to manufacturer: 06-may-2022.Section h-3: device evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed that the lens was received with one haptic bent and twisted.The lens was cleaned, and creases could be found on the optic body, consistent with a lens that was folded.No issues could be identified that would contribute to a delivery issue.The complaint issue of haptic damaged was identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|