The reported event was unconfirmed because the reported failure could not be reproduced.The device did not fail to meet relevant specifications.The product had not caused the reported failure.No root cause could be found because the reported event was unconfirmed.A potential root cause for this failure mode could be due to low slurry calcium concentration causing collapsed/pinched inflation lumen under balloon or along shaft.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "visually inspect the product for any imperfections or surface deterioration prior to use.Recommended inflation capacities: 3cc balloon: use 5cc sterile water.5cc balloon: use 10cc sterile water.15cc balloon: use 20cc sterile water.20cc balloon: use 25cc sterile water.30cc balloon: use 35cc sterile water.40cc balloon: use 45cc sterile water.75cc balloon: use 50cc sterile water.Do not exceeded recommended capacities.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe "stick" in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact and adequately trained professional for assistance, as directed by hospital protocol." corrections: d,h.H11:section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|