ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TORIC UV ABSORBING TRIFOCAL IOL; LENS, MULTIFOCAL Back to Search Results
Model Number TFAT30
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Halo (2227)
Date of Event 04/14/2021
Type of Reportable Event Serious Injury
Event or Problem Description
A nurse reported following an intraocular lens implant procedure, the patient experienced halos and expressed a need for better distance vision over better up close vision, as well as halos.Many interventions were done like drops, and photorefractive keratectomy (prk) (5/20).The lens was later exchanged in a secondary procedure.No patient harm was noticed.Additional information has been requested.
 
Additional Manufacturer Narrative
The lens was returned in a plastic bag with a surgical sponge.Blood and solution was dried on the lens.The optic was cut into pieces, typical of insertion and removal.Power and resolution testing could not be conducted due to the extensive optic damage.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Qualified associated products were indicated.The file indicated the use of a non-qualified viscoelastic.The product investigation could not identify a root cause for the reported complaint.Power and resolution testing could not be conducted due to the extensive optic damage.The power and resolution of each lens is 100% evaluated during the manufacturing process to determine acceptability per model and diopter.Information was provided that the 19.5 diopter, (1.50cyl) add power +2.2 & 3.2 lens was replaced with a 19.5 diopter, (1.5 extended depth of focus) lens.Due to the exchange of a lens to an extended depth of focus lens with a lower diopter may suggest that the replacement lens model may have been an improved choice for the patient's vision needs.No additional information has been provided at this time.There are no other complaints in this lot.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX TORIC UV ABSORBING TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14142111
Report Number1119421-2022-00795
Device Sequence Number4025326
Product Code MFK
UDI-Device Identifier00380652418690
UDI-Public00380652418690
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model NumberTFAT30
Device Catalogue NumberTFAT30.195
Device Lot Number12756641
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2022
Initial Report FDA Received Date04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C; MONARCH III IOL DELIVERY SYSTEM, INJECTOR; NUVISC
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-