• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® BALLOON DILATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® BALLOON DILATION CATHETER Back to Search Results
Model Number 998706
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/06/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the pressure pump was broken and functioned improperly, so water leaks occurred in the ureteral balloon catheter during inflation.Per follow up information received on (b)(6) 2022, the leak occurred on the pressure pump at the connection of pressure pump and catheter.There was no issue with the balloon dilation catheter.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned for evaluation.A potential root cause for this event could be, "inappropriate package design".The device was used for treatment purposes.It was unknown if the device had met all relevant specifications or resulted in the reported event.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "before use, inspect the device to verify that no damage has occurred during shipping and handling." correction: g.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that the pressure pump was broken and functioned improperly, so water leaks occurred in the ureteral balloon catheter during inflation.Per follow up information received on 14apr2022, the leak occurred on the pressure pump at the connection of pressure pump and catheter.There was no issue with the balloon dilation catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® X-FORCE® BALLOON DILATION CATHETER
Common Device Name
DILATION CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14142473
Report Number1018233-2022-02697
Device Sequence Number15010618
Product Code EZN
UDI-Device Identifier10801741052740
UDI-Public(01)10801741052740
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K050875
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2024
Device Model Number998706
Device Catalogue Number998706
Device Lot NumberBMFXFM06
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/24/2022
Supplement Date Received by Manufacturer06/24/2022
Initial Report FDA Received Date04/19/2022
Supplement Report FDA Received Date07/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-