ABBOTT CURRENT PLUS VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD1211-36
Medical Device Problem Code Failure to Charge (1085)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Further information was requested but not received.
 
Event or Problem Description
Patient presented in-clinic after a charge time out occurred on the device.The device was explanted and replaced to resolve the event.The patient was in stable condition.
 
Additional Manufacturer Narrative
The reported field event patient of long charge time was confirmed via a review of device data.Interrogation of the device revealed the device was above elective replacement indicator (eri) when received.Telemetry, impedance, sensing, pacing, and high voltage (hv) output functions of the device were tested and found to be normal.A capacitor maintenance test was successfully performed, and charge time was normal.The cause of the long charge time could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURRENT PLUS VR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14142505
Report Number2017865-2022-08296
Device Sequence Number20411002
Product Code NVZ
UDI-Device Identifier05414734503686
UDI-Public05414734503686
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2017
Device Model NumberCD1211-36
Device Catalogue NumberCD1211-36
Device Lot NumberS000022321
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/07/2022
Supplement Date Received by Manufacturer05/03/2022
07/29/2022
Initial Report FDA Received Date04/19/2022
Supplement Report FDA Received Date05/18/2022
08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-