| Catalog Number |
955607 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Cardiac Arrest (1762)
|
| Date of Event |
03/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The reporter contacted baxter requesting a functional check on an ak98 control unit as a patient coded while performing continuous renal replacement therapy (crrt).Medical intervention and patient outcome were not reported.There was no issue reported with the ak98 control unit and the hospital was only following their policy of having the machine checked out by manufacturing when a patient injury is reported.Despite attempts to obtain additional information related to the patient event, patient¿s demographics, medical condition, or medical history, detailed related to crrt, no additional information was able to be obtained.
|
| |
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: it was reported that the patient coded and passed away and the cause of death was likely due to pulmonary embolism.H10: the device was evaluated on site by a baxter qualified technician which confirmed the device functioned as intended throughout treatment.A one hour simulation was performed with no issues noted.No device malfunction was identified.Based on information from the device evaluation and additional information received from the reporter, the prismax control unit was determined not to be a factor in the event of patient death.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|