Model Number ACE36E
Medical Device Problem Code Failure to Read Input Signal (1581)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/16/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a thyroidectomy the ace36e was not working.The procedure was delayed 10 minutes.It was completed successfully.
 
Additional Manufacturer Narrative
(b)(4).Batch #: p92906.Additional information was requested and the following was obtained: did the generator display any error messages and if so what were they? tighten assembly / identifying instruments did the device pass the pre-test? yes had the device been working and then stopped? yes did the patient experience any adverse consequences due to this issue?- no investigation summary the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.The device was received with the blade discolored (blue) due to heat generated from contact between the blade and tissue pad.However, the blade was not cracked.In addition the tissue pad damaged was noted to be melted and 100% present.The device was connected to a test hand piece and a gen11.During functional testing, it was noted that the hand activation buttons were nonfunctional.However, the device did activate when tested with the foot switch.Probable cause of blade discoloration is applying pressure between the instrument blade and tissue pad without having tissue between them.The device was disassembled to inspect the internal components.Corrosion was found at the hand activation domes.Due to the moisture discovered on the instrument, which is evidence of possible reuse, we are unable to determine how this condition impacted the performance of the device and, therefore, cannot conclude root cause.Our manufacturing, sterilization, packaging, and shipment processes do not introduce corrosion / moisture to the device.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result activation issues such as failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damage blade can result in a broken blade.Probable causes of tissue pad damage are applying pressure between the instrument blade and tissue pad without having tissue between them.Avoid activating the blade without tissue between the blade and tissue pad to prevent damage to the tissue pad.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.As part of ethicon endo surgery quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance's were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE 36CM W ERG HANDLE
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
MDR Report Key14142919
Report Number3005075853-2022-02491
Device Sequence Number4293471
Product Code LFL
UDI-Device Identifier10705036000938
UDI-Public10705036000938
Combination Product (Y/N)N
PMA/510(K) Number
K042777
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model NumberACE36E
Device Catalogue NumberACE36E
Device Lot NumberP92906
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/28/2022
Initial Report FDA Received Date04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/03/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR AND HANDPIECE.
Patient SexUnknown
-
-