• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number MS1-5590R
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event or Problem Description
Information was received that the rod failed to distract.No patient adverse event was reported.
 
Event or Problem Description
No additional information has been provided.
 
Additional Manufacturer Narrative
Additional data: h3, h6, h10 device evaluation: upon return, visual inspection of the returned rod revealed the rod was partially distracted with score marks consistent with expected markings due to incremental distraction.The wear/score marks observed on the distraction rod indicate the distraction rod was extended approximately 4 mm from its initial position.X-ray images of the internal components showed no damage and revealed the rod was partially distracted.The rod was functionally tested and was unable to distract and retract with the manual distracter as well as the external remote controller (erc).The rod was therefore determined to be jammed.The rod was then sectioned and titanium debris build up was found which may have caused the reduced functionality of the rod.Sectioning of the rod determined that the rod could not be separated from the housing without use of high force applied which means possible bending forces had been applied to the rod from patient anatomy/activity which then may have caused the distraction rod to wedge into the housing tube, then causing the rod to jam.Device records review: review of the device history records (dhr) revealed the rod was manufactured in accordance with the specified requirements and met all of the required quality inspections prior to release.
 
Additional Manufacturer Narrative
Corrected data: d4 (model number and udi number) additional data: b5, d9, h3, h6, h10.
 
Event or Problem Description
No additional information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key14145388
Report Number3006179046-2022-00181
Device Sequence Number3986173
Product Code PGN
UDI-Device Identifier00812258026325
UDI-Public812258026325
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K201543
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMS1-5590R
Device Lot NumberA171204-03
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer05/20/2022
06/16/2022
Initial Report FDA Received Date04/19/2022
Supplement Report FDA Received Date05/26/2022
06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/04/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age10 YR
Patient SexPrefer Not To Disclose
Patient Weight45 KG
-
-