MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 68 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 68 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Model Number 68LP-275
Medical Device Problem Codes Fluid/Blood Leak (1250); Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 10/18/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported "deflation and capsular contracture baker grade iii" against right device.The device remains implanted.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii, deflation.
 
Event or Problem Description
The device has been explanted.
 
Event or Problem Description
Healthcare professional reported "deflation and capsular contracture baker grade iii" against right device.The device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified fold creases, wear abrasion, one linear opening, yellow particles in device inner surface.A microscopic analysis and leak test were performed which identified one opening crease smooth.Fill inspection was performed and identified no blockage in the valve.Based on the device analysis the final assessment is: one opening crease smooth in the side posterior assessed as fold flaw opening.Additional, changed, and/or corrected data: h3, h6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 68 SALINE FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
chris sprague
12331-a riata trace parkway
building 3
austin, TX 78727
8479375615
MDR Report Key14146231
Report Number9617229-2022-06097
Device Sequence Number4293580
Product Code FWM
UDI-Device Identifier10888628000575
UDI-Public10888628000575
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P990074
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/24/2022
Device Model Number68LP-275
Device Catalogue Number68LP-275
Device Lot Number3181555
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer04/21/2022
05/17/2022
Initial Report FDA Received Date04/19/2022
Supplement Report FDA Received Date05/13/2022
06/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/24/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age40 YR
Patient SexFemale
Patient Weight80 KG
-
-