MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN POROUS CR FEM SZ 7R; PRSTHSS, KNEE,PAT/FMROTIBIAL,SEMI-CNSTRND,UNCMNTD,POROUS,COATED,POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN POROUS CR FEM SZ 7R; PRSTHSS, KNEE,PAT/FMROTIBIAL,SEMI-CNSTRND,UNCMNTD,POROUS,COATED,POLY/METAL/POLY Back to Search Results
Model Number 71423257
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/11/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after a tka performed on (b)(6) 2022, the patient experience an infection after 6 weeks.The adverse event was addressed via revision surgery on (b)(6) 2022 to explant the femur, tibia, insert and patellar components.Patient's current status of health is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, this case reports a revision surgery due to infection approximately six weeks after tka.No further clinically relevant information was provided for evaluation.Without clinically relevant information for evaluation, the root cause of the reported infection cannot be determined.The patient impact beyond the reported revision could not be assessed.No further medical assessment could be rendered at this time.If additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed that acute post-surgical wound infection and late deep wound sepsis has been identified as a possible adverse effects.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the sterilization records revealed the batch was sterilized within normal parameters.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but are not limited to contamination, patient reaction, and post-operative healing issue.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN POROUS CR FEM SZ 7R
Common Device Name
PRSTHSS, KNEE,PAT/FMROTIBIAL,SEMI-CNSTRND,UNCMNTD,POROUS,COATED,POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
MDR Report Key14147789
Report Number1020279-2022-01864
Device Sequence Number12245454
Product Code MBH
UDI-Device Identifier00885556030769
UDI-Public00885556030769
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K073325
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model Number71423257
Device Catalogue Number71423257
Device Lot Number20FHA0001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2022
Supplement Date Received by Manufacturer03/28/2022
05/19/2022
Initial Report FDA Received Date04/19/2022
Supplement Report FDA Received Date04/20/2022
05/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexMale
-
-