| Model Number |
GLU TEST STRIP |
| Medical Device Problem Code |
Imprecise or Erratic Results (4059)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer indicated that their stat strip meter, serial number: (b)(4) gave a lower result compared to another meter.Patient was transferred to pediatrics and tested again.Customer confirmed that the stat strip meter was passing qc.Using test strip lot 0320153249 pn 54515.It was indicated that there was no adverse event due to the meter.
|
| |
|
Additional Manufacturer Narrative
|
|
There is currently a pending investigation.Nova is requesting additional information, and further details will be provided in a supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi: (b)(4).The complainants alleged deficiency could not be confirmed.The statstrip glucose test strips in question were not returned for evaluation.An investigation was performed using qa retain strips from the same lot.The retain strips met all test specifications.The dhr was complete and contained all relevant data indicating the product put into finished goods met all specifications.Nova biomedical will continue to monitor for recurrence as part of our post market surveillance program.
|
| |
|