BOSTON SCIENTIFIC CORPORATION FLEXIVA PULSE ID; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number M006L8406931
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has been received for analysis; however, the analysis has not yet been completed.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022 that a flexiva pulse id laser fiber was used in a bladder stone procedure performed on (b)(6) 2022.The laser unit used was a lumenis 100 watt laser console.During preparation, the physician noticed that the laser fiber jacket cover extended to the very end of the laser fiber.The procedure was completed with another flexiva pulse id laser fiber.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: medical device problem code a0501 captures the reportable event of fiber tip detached.Block h10: investigation results the returned flexiva pulse id laser fiber was analyzed, and a visual evaluation noted that it was returned in one piece.The laser fiber measured 278.9 cm from the tip of the connector to the end of the exposed glass tip.The exposed glass was detached.A functional evaluation revealed that the light from the sma connector was bright and round, indicating that there was no fracture within the fiber connector.No other problems with the device were noted.The reported event of the flexiva pulse id laser fiber tip detached was confirmed.It is likely that procedural conditions and handling of the device during use contributed to the fiber tip damage.Therefore, the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022 that a flexiva pulse id laser fiber was used in a bladder stone procedure performed on (b)(6) 2022.The laser unit used was a lumenis 100 watt laser console.During preparation, the physician noticed that the laser fiber jacket cover extended to the very end of the laser fiber.The procedure was completed with another flexiva pulse id laser fiber.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIVA PULSE ID
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14147917
Report Number3005099803-2022-02072
Device Sequence Number16880333
Product Code GEX
UDI-Device Identifier08714729978749
UDI-Public08714729978749
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K210925
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/20/2024
Device Model NumberM006L8406931
Device Catalogue Number2130-02
Device Lot Number0028241074
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer04/26/2022
Initial Report FDA Received Date04/19/2022
Supplement Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-