Model Number M00547000
Medical Device Problem Codes Inflation Problem (1310); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has been received for analysis; however, the analysis has not yet been completed.Upon completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that an extractor pro xl retrieval balloon was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, the physician asked to put the balloon to 9mm; however, the balloon was not inflated when the balloon was pulled out of duct.Prior to reinserting back into duct the balloon was inflated and deflated a couple of times, the balloon did inflate but did not stay up at 9mm, only 12mm.Additionally, the balloon shape and size was noted to be odd and not rounded, the one side was under-inflated and did not look like it was inflated all the way to 12mm.The balloon was then removed from the duct, the stopcock was opened but the balloon did not automatically deflate and the syringe plunger had to be physically pulled back.The procedure was completed with another extractor pro xl retrieval balloon.There were no patient complications reported as a result of this event.
 
Event or Problem Description
It was reported to boston scientific corporation that an extractor pro xl retrieval balloon was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, the physician asked to put the balloon to 9mm; however, the balloon was not inflated when the balloon was pulled out of duct.Prior to reinserting back into duct the balloon was inflated and deflated a couple of times, the balloon did inflate but did not stay up at 9mm, only 12mm.Additionally, the balloon shape and size was noted to be odd and not rounded, the one side was under-inflated and did not look like it was inflated all the way to 12mm.The balloon was then removed from the duct, the stopcock was opened but the balloon did not automatically deflate and the syringe plunger had to be physically pulled back.The procedure was completed with another extractor pro xl retrieval balloon.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: device code a0406 captures the reportable event of the balloon irregular shape.Block h10: investigation results a visual examination of the returned complaint device found that the balloon and the catheter of the device had no damages, which does not confirm the reported event of balloon irregular shape.Microscopic inspection was done and found the guidewire channel of the device was ripped.Functional evaluation was performed, the device was inflated with air, the balloon was inflated without problem, no damages were found and the device works properly.The returned device review showed no evidence of either the alleged (or any problem which could have contributed to the event.Therefore, the most probable root cause for this complaint is no problem detected.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.A labeling review was performed and from the information available, this device was used per the instructions for use (ifu)/product label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTRACTOR PRO RX
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14147932
Report Number3005099803-2022-02106
Device Sequence Number2946165
Product Code FGE
UDI-Device Identifier08714729790266
UDI-Public08714729790266
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K102082
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/17/2023
Device Model NumberM00547000
Device Catalogue Number4700
Device Lot Number0028576291
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/24/2022
Supplement Date Received by Manufacturer04/21/2022
Initial Report FDA Received Date04/19/2022
Supplement Report FDA Received Date05/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-