DATEX-OHMEDA, INC. AESPIRE 7100; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Inaccurate Delivery (2339)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/25/2022
Event Type  malfunction  
Event Description
The hospital reported a malfunction that could cause a hypoxic mixture in the patient circuit.There was no report of patient involvement.
 
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The n2o (nitrous oxide) mixer valve was replaced to correct the issue.No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Date of device manufacture year is 2003.The month of manufacture was unavailable at time of mdr filing.Legal manufacturer: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESPIRE 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
anthony amenson
MDR Report Key14148056
MDR Text Key289732802
Report Number2112667-2022-00932
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/25/2022
Initial Date FDA Received04/19/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-