• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC LIVER ACCESS AND BIOPSY SET; FCG KIT, NEEDLE, BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC LIVER ACCESS AND BIOPSY SET; FCG KIT, NEEDLE, BIOPSY Back to Search Results
Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) #: preamendment.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that a patient required a liver access and biopsy set for a transjugular liver biopsy procedure.During the procedure, the proximal hub separated from the body of the catheter due to "manipulation" of the device.No portion of the device remained inside the patient's body.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.Additional information regarding the event and patient outcome has been requested but is currently unavailable.
 
Additional Manufacturer Narrative
This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Additional information was received on 03may2022 clarifying that "manipulation" meant "rotate the catheter" while engaging the catheter into the hepatic vein.No tortuous patient anatomy present.The procedure was successfully completed by using another similar device.
 
Event or Problem Description
No additional information regarding patient and/or event details has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged or unavailable.Investigation ¿ evaluation.(b)(6) hospital informed cook on 04apr2022 of an issue with a liver access and biopsy set (rpn: labs-100-j, lot: 10339909).The customer stated the proximal hub of the 5fr straight catheter was disengaged from the body due to manipulation (catheter rotation) during the procedure.A replacement set was used to complete the procedure successfully.Reviews of the documentation, including the complaint history, device history record, instructions for use (ifu), manufacturing instructions and quality control procedures of the device, were conducted during the investigation.The complaint device was not returned for evaluation; therefore, no physical examinations could be conducted.However, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) for lot 10339909 found no nonconformances that could have contributed to the reported failure mode.It should be noted that there were no other complaints associated with the final product lot number.Cook also reviewed product labeling.The product ifu, [t_labs_rev6] ¿liver access and biopsy set,¿ provides the following information to the user related to the reported failure mode: instructions for use.Utilizing standard access techniques, introduce an appropriate.035 inch wire guide into the inferior vena cava via jugular vein access.Using a selective catheter and wire guide of choice, catheterize the right hepatic vein or an appropriate alternative hepatic vein branch.Leave the wire in a safe, distal position, and remove the catheter.Caution: to prevent cardiac arrhythmias, continuous cardiac monitoring is recommended while negotiating past the right atrium.Introduce and advance the transjugular introducer sheath and stiffening cannula over the wire guide into the selected hepatic vein.When introducing these components as a preassembled set, the included straight catheter may be used to facilitate introduction.Once the set is positioned within the hepatic vein, the straight catheter should be removed.Note: care should be taken to prevent damage to the straight catheter during removal through the metal stiffening cannula.Leaving a wire guide through the straight catheter during removal may aid in preventing catheter damage.How supplied.Supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened or undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry, cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred.¿ upon review of the dmr and dhr, there is no indication the complaint device was manufactured out of specification.There is no evidence of nonconforming material in house or in the field.Based on the information provided, no returned product, and the results of our investigation, it was concluded that an adverse event related to the procedure contributed to the event.The customer stated the proximal hub of the 5fr straight catheter was disengaged from the body due manipulation (catheter rotation) in the procedure.The appropriate personnel have been notified.Per the risk assessment no further action is required.We will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIVER ACCESS AND BIOPSY SET
Common Device Name
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14148136
Report Number1820334-2022-00614
Device Sequence Number17581365
Product Code FCG
UDI-Device Identifier00827002507791
UDI-Public(01)00827002507791(17)230219(10)10339909
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/19/2023
Device Model NumberN/A
Device Catalogue NumberLABS-100-J
Device Lot Number10339909
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/04/2022
Supplement Date Received by Manufacturer05/03/2022
11/28/2022
Initial Report FDA Received Date04/19/2022
Supplement Report FDA Received Date05/06/2022
12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/19/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-