Model Number U228
Device Problem High impedance (1291)
Patient Problem Device Overstimulation of Tissue (1991)
Event Date 02/16/2022
Event Type  malfunction  
Event Description
It was reported that this system with a left ventricular (lv) lead and a cardiac resynchronization therapy pacemaker (crt-p) showed stimulation the night of the day it was implanted.A lead safety switch (lss) was triggered.Dutring the implant, pacing impedances were high, within normal limits.The device was reprogrammed to prevent (lss) from happening.The lv lead and the crt-p remain in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONIST X4 CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14152727
MDR Text Key289691910
Report Number2124215-2022-12116
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559471
UDI-Public00802526559471
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S138
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/10/2024
Device Model NumberU228
Device Catalogue NumberU228
Device Lot Number769382
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/16/2022
Initial Date FDA Received04/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/24/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age84 YR
-
-