• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INNOVA VASCULAR; STENT, SUPERFICIAL FEMORAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INNOVA VASCULAR; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Model Number 26924
Medical Device Problem Codes Fracture (1260); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Initial reporter facility name: (b)(6).
 
Event or Problem Description
It was reported that the stent fractured.A 6x60x130 innova self expanding stent was selected for use.The target lesion was located in the common femoral artery.The stent was successfully placed.A non-bsc device was used to close the vascular access site.When it was placed, it pushed on the stent and caused the stent to fracture.The patient had to be repunctured, and an additional innova stent was successfully placed within the original stent to make it durable in the vessel.The procedure was successfully completed.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA VASCULAR
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14153662
Report Number2134265-2022-03954
Device Sequence Number5932021
Product Code NIP
UDI-Device Identifier08714729874003
UDI-Public08714729874003
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P140028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/03/2026
Device Model Number26924
Device Catalogue Number26924
Device Lot Number0027775172
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/31/2022
Initial Report FDA Received Date04/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VERYAN CELT VASCULAR CLOSURE DEVICE
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-