SMITH & NEPHEW MEDICAL LTD. ALLEVYN TRACHEOSTOMY 3.5X3.5" CTN 10; DRESSING,WOUND,OCCLUSIVE Back to Search Results
Model Number 66027640
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Code Hypersensitivity/Allergic reaction (1907)
Date of Event 03/24/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during treatment the patient had a severe allergic reaction to allevyn trach dressing.This adverse event was resolved by supplying epinephrine x2 and benadryl x1 and removing the dressing.Current health status of the patient is ok.
 
Additional Manufacturer Narrative
Additional information: d4 (expiration date), d8, h4, h6 (component code, type of investigation, investigation findings, investigation conclusions).Results of investigation: the device, used in treatment, was not returned for evaluation.All provided information has been reviewed and we have not been able to confirm a relationship between the device and the reported event or determine a root cause.Probable cause include incorrect skin preparation, incorrect dressing application or the patient experienced a reaction to one or more of the components of the dressing.The users of the reported product are advised to consult the ifu, to prevent future occurrences of the reported issue.This guide provides comprehensive instructions of the operation, use and limitations of the device, including the correct skin preparation technique and dressing application.The batch record was reviewed and it could be confirmed that the products had progressed through a satisfactory release process which involved testing the product against the requirements of the finished product specification.There was also no issue identified during the manufacturing process that could have caused or contributed to the complaint problem.A review of complaint history did not reveal similar events for this nature with no manufacturing problems observed.A clinical/medical assessment was performed and determined no further actions are required.A risk management review concluded that the alleged failure mode and any associated harm has been anticipated within the risk file, with no updates required.A review of prior escalation actions found no actions applicable to the scope of this case.This investigation is now complete with no further action deemed necessary.Smith and nephew will continue to monitor for any adverse trends relating to this product range.Corrected data: b1, d1, d4 (catalog number & udi), h6 (health effect - impact code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEVYN TRACHEOSTOMY 3.5X3.5" CTN 10
Common Device Name
DRESSING,WOUND,OCCLUSIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14153877
Report Number8043484-2022-00123
Device Sequence Number2944112
Product Code NAD
UDI-Device Identifier05000223415860
UDI-Public05000223415860
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2025
Device Model Number66027640
Device Catalogue Number66027640
Device Lot Number202206
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2022
Supplement Date Received by Manufacturer10/12/2022
Initial Report FDA Received Date04/19/2022
Supplement Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age12 YR
Patient SexMale
Patient Weight25 KG
-
-