ANGIOMED GMBH & CO. MEDIZINTECHNIK KG VENOVO VENOUS STENT Back to Search Results
Model Number VENUM16060
Medical Device Problem Codes Migration or Expulsion of Device (1395); Difficult to Remove (1528); Obstruction of Flow (2423)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 08/23/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that approximately ten days post stent placement procedure, the device was allegedly migrated from the common iliac artery to the vena cava.It was further reported that a blood clot was identified and removed.Reportedly, the device was unable to be removed.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Expiry date: 05/2021.
 
Additional Manufacturer Narrative
H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was not returned to the manufacturer for evaluation and images were not provided for review.Based on the information available the investigation is inconclusive result for reported issues.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.A stent size selection table is part of the instructions for use describing the relationship between reference vessel diameter and unconstrained stent inner diameter.Regarding pta the instructions for use states: "predilation of chronic lesions with a balloon dilatation catheter is recommended.If performed, select a balloon catheter that matches the size of the reference vessel.", and "post stent expansion with a balloon dilatation catheter is recommended." the instructions for use closely describes holding and handling of the system throughout deployment.Stent migration was found mentioned as a potential adverse event that may occur.H10: d4(expiry date: 05/2021), g3.H11: g1, h6 (method).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that approximately ten days post stent placement procedure, the device was allegedly migrated from the common iliac artery to the vena cava.It was further reported that a blood clot was identified and removed.Reportedly, the device was unable to be removed.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENOVO VENOUS STENT
Common Device Name
VENOUS STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14155628
Report Number9681442-2022-00117
Device Sequence Number20316296
Product Code QAN
UDI-Device Identifier00801741103568
UDI-Public(01)00801741103568
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P180037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVENUM16060
Device Catalogue NumberVENUM16060
Device Lot NumberANDS1277
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/24/2022
Supplement Date Received by Manufacturer05/13/2022
Initial Report FDA Received Date04/20/2022
Supplement Report FDA Received Date05/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age36 YR
Patient SexMale
-
-