• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-028
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported a 28 mm amplatzer septal occluder was chosen for procedure on (b)(6)2022.During deployment of the occluder, the device presented with a deformation and the device was not placed.The patient remained stable throughout the procedure.The patient was noted to have been discharged.There was no clinically significant delay in procedure and no adverse patient effects.
 
Additional Manufacturer Narrative
An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional and dimensional specifications when analyzed under non-physiological conditions.A broken wire was noted on the distal disc.Based on the information received, the cause of the reported incident could not be conclusively determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key14156021
Report Number2135147-2022-00179
Device Sequence Number17544989
Product Code MLV
UDI-Device Identifier05415067019325
UDI-Public05415067019325
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model Number9-ASD-028
Device Catalogue Number9-ASD-028
Device Lot Number7281654
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/24/2022
Supplement Date Received by Manufacturer08/09/2022
Initial Report FDA Received Date04/20/2022
Supplement Report FDA Received Date09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/20/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AMPLATZER TORQVUE DS 9-ITV10F45/80 LOT: 6444362
Patient SexUnknown
-
-