MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA BALL HEADS: BIPOLAR HEAD BIPOLAR HEAD Ø28X47; HIP ACETABULAR BIPOLAR HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA BALL HEADS: BIPOLAR HEAD BIPOLAR HEAD Ø28X47; HIP ACETABULAR BIPOLAR HEAD Back to Search Results
Model Number 25060.2847
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/24/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Clinical evaluation performed by medical affairs department intraprosthetic dislocation 1 month after the partial hip prosthesis surgery in a 85 year old patient.Dislocation of the head from the bipolar cup is normally prevented by the retention mechanism.Technical analysis of the implant might reveal if a defect was present in said mechanism; to date we have no evidence of such situation.Conditions that may cause intra prosthetic dislocation in a well functioning bipolar cup are difficult to reproduce.With information at hand we are unable to draw any definitive conclusion.Batch review performed on (b)(6) 2022: lot 2103619: 12 items manufactured and released on (b)(6) 2021.Expiration date: (b)(6) 2026.No anomalies found related to the problem.To date, 4 items of the same lot have been sold without any similar reported event.Other device involved: ball heads: cocr 01.25.013 cocr ball head 12/14 ø 28 size l +3.5, (k072857), lot.2107132: 100 items manufactured and released on (b)(6) 2021.Expiration date: (b)(6) 2026.No anomalies found related to the problem.To date, 25 items of the same lot have been sold without any similar reported event.
 
Event or Problem Description
Luxation of the bipolar head from the metallic head (intra-prosthetic dislocation).The surgeon revised all components.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALL HEADS: BIPOLAR HEAD BIPOLAR HEAD Ø28X47
Common Device Name
HIP ACETABULAR BIPOLAR HEAD
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key14157400
Report Number3005180920-2022-00272
Device Sequence Number2945343
Product Code KWY
UDI-Device Identifier07630030843655
UDI-Public07630030843655
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K091967
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/19/2026
Device Model Number25060.2847
Device Catalogue Number25060.2847
Device Lot Number2103619
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2022
Initial Report FDA Received Date04/20/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age85 YR
Patient SexMale
Patient RaceWhite
-
-