Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/22/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Ge healthcare's investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Unique identifier: (b)(6).Legal manufacturer: (b)(4).Device evaluation anticipated, but not yet begun.
|
|
Event Description
|
The hospital reported a leak in excess of 4.5lpm.There was no report of patient involvement.
|
|
Manufacturer Narrative
|
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The o2 cell, cooling fan, and upper table of the ventilator block were replaced to resolve the issue.
|
|