ENCORE MEDICAL L.P. X-ALT HIP; LINER, ACETABULAR, HOODED, NEUTRAL, 10°, HXL/FMP, 32MM ID, MP6
|
Back to Search Results |
|
| Model Number |
932-32-250 |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
|
| |
|
Event or Problem Description
|
|
Revision surgery - patient complaint of hip clicking.Poly liner was worn through at the superior-posterior 1/3 margin of the liner.The cup was originally placed in a very vertical position as indicated on x-ray.Metal head was also removed with some minor scuffing of the surface.
|
| |
|
Additional Manufacturer Narrative
|
|
The reason for this revision surgery was reported as worn poly.The actual length of in-vivo for the items listed is unknown as the original surgery date was not provided or could be established.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.This investigation is limited in scope as only partial information was provided djo surgical for review.The revised items were not returned for examination and the lot numbers were not provided.To adequately investigate this event, the part and lot numbers are necessary.Given the limited information, a search for an invoice (of the previous surgery) produced no results, therefore; the items removed could not be verified.If this information is submitted at a future date, this investigation will be re-evaluated.Customer complaint history of the reported items showed no present trends or on-going issues that are needing a review.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.The surgeon performed this revision to remedy the patient's condition.This complaint will be closed pending receipt of additional information.No further action is deemed necessary at this time.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
|
| |
|
Search Alerts/Recalls
|
|
|