MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® SILICONE URETERO-INTESTINAL ANASTOMOSIS URINARY DIVERSION STENT; URINARY STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® SILICONE URETERO-INTESTINAL ANASTOMOSIS URINARY DIVERSION STENT; URINARY STENT Back to Search Results
Model Number 089007
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Post Operative Wound Infection (2446); Unspecified Kidney or Urinary Problem (4503)
Date of Event 03/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the survey respondent mentioned about surgical site infection and irritative urinary symptoms when asked about the complications faced while using the urinary stent.Medical intervention was unknown.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the survey respondent mentioned about surgical site infection and irritative urinary symptoms when asked about the complications faced while using the urinary stent.Medical intervention was unknown.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be ¿material selection".The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "a variety of techniques may be used during stent placement.Each surgeon should use the methods with which he/she is familiar.1.During the surgical procedure, after the ureter has been divided, the stent is straightened by inserting the guidewire into the proximal end (open end) of the stent and advancing it forward to distal end (closed end).2.Pass the stent through the open end of the cut ureter up into the renal pelvis.3.While holding the proximal end, withdraw the guidewire.This will allow the ¿figure 4¿ to form within the renal pelvis thereby ensuring proper placement.4.After isolation of the bowel segment and closure of the distal end have been accomplished, pass a long right angle clamp through the stomal end to the point where the ureter is to be anastomosed.5.Draw the proximal end of the urinary diversion stent through the dissection to the stomal end.6.Palpate the renal pelvis to ensure proper placement of the distal end.7.After the external stoma has been formed, trim the stent so that approximately 5cm protrudes from the stoma.Then attach the stent to the skin with a nonabsorbable suture." the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® SILICONE URETERO-INTESTINAL ANASTOMOSIS URINARY DIVERSION STENT
Common Device Name
URINARY STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key14160975
Report Number1018233-2022-02726
Device Sequence Number2945540
Product Code FAD
UDI-Device Identifier10801741011648
UDI-Public(01)10801741011648
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K823487
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number089007
Device Catalogue Number89007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer06/02/2022
Initial Report FDA Received Date04/20/2022
Supplement Report FDA Received Date06/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-