• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL CADD MEDICATION CASSETTES; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL CADD MEDICATION CASSETTES; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7302-24
Device Problem Connection Problem (2900)
Patient Problem Inflammation (1932)
Event Date 02/28/2022
Event Type  malfunction  
Manufacturer Narrative
No product or photographic evidence were provided to aid in this investigation.The complaint report offered insufficient details to determine whether this product functioned as intended, or was used in a manner consistent with its instructions for use (ifu) or failed to meet product specifications.Lacking any additional evidence, this complaint has been closed as unconfirmed.If the product is returned, smiths medical will reopen this complaint for further investigation.A manufacturing dhr review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.
 
Event Description
It was reported that both legacy pumps were alarming no disposable alarms.Patient has mixed 2 new cassettes, with same results.Patient has troubleshot with no resolution.Patient reports previous issue with pump that caused him to receive more medication than programmed, states feet, mouth, hands felt weird.Patient able to resolve alarm with 3rd cassette mixed.2 affected cassette lot numbers: 4196398, expire: 2026-09-21.Patient off medication for 1 hour, asymptomatic.Patient will waste 3 partial cassettes, states has enough medication to last until next order.No further details provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD MEDICATION CASSETTES
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14161044
MDR Text Key289662480
Report Number3012307300-2022-06720
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586027239
UDI-Public10610586027239
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 04/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number21-7302-24
Device Catalogue Number21-7302-24
Device Lot Number4196398
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-