• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP 3.0MM CARBIDE MATCH HEAD; DRILLS, BURRS, TREPHINES & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP 3.0MM CARBIDE MATCH HEAD; DRILLS, BURRS, TREPHINES & ACCESSORIES Back to Search Results
Model Number 5820107530C
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow up report will be filed once the quality investigation is complete.Awaiting device return.
 
Event or Problem Description
It was reported that during a thoracolumbar spine revision/fusion, the bur broke in multiple locations.It was also reported that there were no adverse consequences and no delays as a result of this event.It was further reported that the procedure was completed successfully.
 
Event or Problem Description
It was reported that during a thoracolumbar spine revision/fusion, the bur broke in multiple locations.It was also reported that there were no adverse consequences and no delays as a result of this event.It was further reported that the procedure was completed successfully.
 
Additional Manufacturer Narrative
H3 & h6: investigation results indicate that the bur broke due to issues with the associated elite 14cm angled attachment 5407120482 sn (b)(6).H6: the quality investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.0MM CARBIDE MATCH HEAD
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES
Manufacturer (Section D)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer (Section G)
STRYKER INSTRUMENTS-IRELAND
instruments division
carrigtwohill bus. & tech park
carrigtwohill NA
EI   NA
Manufacturer Contact
una barry
instruments division
carrigtwohill bus. & tech park
carrigtwohill NA
EI   NA
214532900
MDR Report Key14161738
Report Number3015967359-2022-00701
Device Sequence Number5755867
Product Code HBE
UDI-Device Identifier07613327294989
UDI-Public07613327294989
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K143320
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5820107530C
Device Catalogue Number5820107530C
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/29/2022
Supplement Date Received by Manufacturer06/09/2022
Initial Report FDA Received Date04/20/2022
Supplement Report FDA Received Date06/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5407120482 SN(B)(6)
Patient SexUnknown
-
-