Model Number 30914
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that the bd alaris¿ extension set leaked from the tubing during use.The following information was provided by the initial reporter: "rn hung medication, returned to check if it was finished and noticed the tubing was wet, floor had puddle and it seemed like the tubing itself was leaking.Minor effect to patient.".
 
Event or Problem Description
It was reported that the bd alaris¿ extension set leaked from the tubing during use.The following information was provided by the initial reporter: "rn hung medication, returned to check if it was finished and noticed the tubing was wet, floor had puddle and it seemed like the tubing itself was leaking.Minor effect to patient.".
 
Additional Manufacturer Narrative
Investigation summary a 30914 extension tubing, a non bd connector and a 50ml bd syringe were received and tested by our quality team.When saline solution was injected into the system via syringe, a leak was noticed between the non bd connector and the extension tubing.This leakage verified the customer's complaint.When the non bd connector was removed however, there was no leakage realized and it is unlikely that there is any issue with the extension tubing.Multiple other connectors were used with the tubing and none leaked.This is evidence that there is not an issue with the set reported in this complaint.A device history record review could not be performed because a lot number was not provided by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS¿ EXTENSION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14162127
Report Number9616066-2022-00443
Device Sequence Number5932352
Product Code FPA
UDI-Device Identifier50885403233317
UDI-Public50885403233317
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K790108
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number30914
Device Catalogue Number30914
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2022
Supplement Date Received by Manufacturer08/10/2022
Initial Report FDA Received Date04/20/2022
Supplement Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-