| Model Number |
CD1257-40 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted.The patient condition was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is included in the battery performance alert advisory issued by abbott on (b)(6) 2017.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is subject the battery performance advisory.Device was received above elective replacement indicator (eri) when interrogated.There was no detection of battery performance alert (bpa).A longevity calculation was normal.Based on this information, there was no evidence of premature depletion.
|
| |
|