• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X45 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X45 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 1896-5045S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 09/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The event could be confirmed, since the screw was found broken during the investigation.The device inspection revealed the following: the locking screw was returned without allegation against the device.However, the visual inspection has shown that the locking screw is broken apart, the fragment with a length of about 20mm was not returned for evaluation.The typical characteristics of a fatigue fracture could be identified at the fracture face.The origin of the crack was at the right corner, there is a fatigue zone with a smooth surface and progression lines over almost the whole surface of the fracture and then we have a shear lip on the left side, where the final fracture occurred.The relevant dimensions were verified during the investigation and no deviation from the specification could be detected.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.The locking screw was returned for evaluation and no product related issue could be detected during the investigation.There was no additional information about the event provided, no x-rays and no information about the patient was made available.Therefore the root cause of the breakage cannot be defined.The evaluation of the received screw has shown that fatigue did lead to the breakage of the device.In this relation following statement from the instructions for use (non-active implant) can be pointed out: "these devices can break when subjected to the increased loading associated with delayed unions and/or non-unions.Internal fixation devices are load sharing devices which are intended to hold fractured bone surfaces in apposition to facilitate healing.If healing is delayed or does not occur, the appliance may eventually break due to metal fatigue.Loads on the device produced by load bearing and the patient¿s activity level will dictate the longevity of the device." if more information is provided, the case will be reassessed.
 
Event or Problem Description
A gamma nail broke at lag screw hole 4 months post-op.During the inspection one of the returned locking screws was found broken due to fatigue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X45 MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14163587
Report Number0009610622-2022-00141
Device Sequence Number5755248
Product Code HSB
UDI-Device Identifier04546540202505
UDI-Public04546540202505
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K032244
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/31/2025
Device Model Number1896-5045S
Device Catalogue Number18965045S
Device Lot NumberK07873B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/24/2022
Initial Report FDA Received Date04/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-