MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII PS INSERT SZ 5-6 15MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII PS INSERT SZ 5-6 15MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 71420836
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Failure of Implant (1924); Discomfort (2330)
Date of Event 03/29/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, after a tka on an unknown date , the patient described a clunking when moving his knee.A revision surgery was performed on (b)(6) 2022 and was discovered the post of the old gii ps insert sz 5-6 15mm had broken off.The current health status of patient is unknown.
 
Additional Manufacturer Narrative
The device was not returned for evaluation but the pictures were reviewed, and the failure mode was confirmed.The clinical/medical investigation concluded that the two provided photos were reviewed and support the reported event but do not aid in the medical investigation of the root cause of the post breakage.No other medical/clinical documentation was provided.Without supporting medical documentation, a thorough medical assessment cannot be performed.In the event additional medical/clinical records are received, the clinical task may be re-opened, and a thorough assessment will be rendered at that time.All documents and images provided as of this date have been reviewed and considered and unless noted do not contribute to the clinical/medical investigation.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed that the patient should be warned that the implant can break or become damaged as a result of strenuous activity or trauma, and has a finite expected service life and may need to be replaced in the future.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to traumatic injury or surgical technique.The contribution of the device to the reported incident could be corroborated as the device was damaged and it was required a revision surgery to treat this event.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS INSERT SZ 5-6 15MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14164054
Report Number1020279-2022-01885
Device Sequence Number20272612
Product Code JWH
UDI-Device Identifier03596010209672
UDI-Public03596010209672
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71420836
Device Catalogue Number71420836
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2022
Supplement Date Received by Manufacturer05/19/2022
Initial Report FDA Received Date04/20/2022
Supplement Report FDA Received Date05/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKN GENESIS II TOTAL KNEE FEM COMP; UNKN GENESIS II TOTAL KNEE TIB BASEPLATE
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-