• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS VERSASAFE EXTENSION SET; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS VERSASAFE EXTENSION SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 10796814
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device expiration date: unknown.Initial reporter facility name: (b)(6) hospital.Fda notified: the initial reporter notified the fda via uf/importer report#: (b)(4).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that the bd alaris versasafe extension set experienced component separation.The following information was provided by the initial reporter: rn noticed blood backing up in the tubing and that the new kind of iv tubing we are using broke.The disconnection was not at a site that normally comes apart.It was where a needless hub and the tubing meet.The tubing slipped apart.
 
Additional Manufacturer Narrative
H6: investigation summary: no photo or sample was received for the customer's complaint of leakage.Without further information like a physical sample for investigation, the complaint cannot be verified and root cause of this failure remains unknown.A device history record review could not be performed because a lot number was not provided by the customer.H3 other text : see h10.
 
Event or Problem Description
It was reported that the bd alaris versasafe extension set experienced component separation.The following information was provided by the initial reporter: rn noticed blood backing up in the tubing and that the new kind of iv tubing we are using broke.The disconnection was not at a site that normally comes apart.It was where a needless hub and the tubing meet.The tubing slipped apart.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS VERSASAFE EXTENSION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14164119
Report Number9616066-2022-00452
Device Sequence Number5968821
Product Code FPA
UDI-Device Identifier50885403234338
UDI-Public50885403234338
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K790108
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10796814
Device Catalogue Number10796814
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2022
Supplement Date Received by Manufacturer07/29/2022
Initial Report FDA Received Date04/20/2022
Supplement Report FDA Received Date08/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-