• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR DRAGONFLY¿ DUO IMAGING CATHETER; INTRAVASCULAR IMAGING CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR DRAGONFLY¿ DUO IMAGING CATHETER; INTRAVASCULAR IMAGING CATHETER Back to Search Results
Model Number DUO
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not returning.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported the procedure was performed in the left anterior descending artery that was mildly tortuous and calcified.A dragonfly duo imaging catheter was inserted, but in was observed on x-ray that a foreign body was on the tip of the device.The catheter was pulled back and it was observed the foreign body moved with the catheter.Therefore, the device was removed.No foreign body was observed in the anatomy on imaging after the catheter was replaced to continue the procedure.There was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no other similar complaints reported from this lot.Based on the information provided, a definitive cause for the foreign body on the tip of the catheter could not be determined.It may be possible that damage occurred to the tip and/or guidewire exit port during advancement giving the appearance of foreign material on the tip while observed on x-ray; however, without having the device to examine, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.D9: device available for evaluation updated from yes to no.
 
Additional Manufacturer Narrative
Visual analysis was performed on the returned device.The reported foreign body on the tip of the device was not confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no other similar complaints reported from this lot.Based on the information provided, a definitive cause for the foreign body on the tip of the catheter could not be determined.Returned analysis was unable to confirm the reported foreign body on the tip of the catheter.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.D9 - device available for eval updated from 'no' to 'yes'.H6 - codes 4114 and 3221 removed.Codes 10 and 213 added.H10 - additional mfg narrative: revised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRAGONFLY¿ DUO IMAGING CATHETER
Common Device Name
INTRAVASCULAR IMAGING CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT MEDICAL REG#3009600098
4 robbins rd
westford MA 01886
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14164122
Report Number2024168-2022-04280
Device Sequence Number5968823
Product Code NQQ
UDI-Device Identifier00183739000708
UDI-Public00183739000708
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K093857
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/14/2023
Device Model NumberDUO
Device Catalogue NumberC408644
Device Lot Number7928178
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/30/2022
Supplement Date Received by Manufacturer05/04/2022
09/12/2022
Initial Report FDA Received Date04/20/2022
Supplement Report FDA Received Date05/16/2022
09/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-