OLYMPUS MEDICAL SYSTEMS CORP. AUTOCLAVABLE CAMERA HEAD Back to Search Results
Model Number OTV-S7H-NA
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device was returned to an olympus service center for evaluation.Upon inspection and testing of the returned device, the reported image loss could not be confirmed or reproduced.However, service found the electrical contacts of the video connector were corroded.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device met all specifications at the time of shipment.It has been over 8 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, the root cause was not able to be determined, since the customer's report could not be duplicated.Per the legal manufacturer, the corrosion on the electrical contacts of the video connector has no potential to cause or contribute to death or serious injury if the malfunction was to recur.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The user facility reported that when preparing the subject device for use during an unspecified procedure, there was no image due to a "sandstorm" on the display.There was no patient or user injury reported due to the event.Additional details have been requested regarding the reported event.However, no additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCLAVABLE CAMERA HEAD
Common Device Name
CAMERA HEAD
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14164621
Report Number8010047-2022-06745
Device Sequence Number4988361
Product Code FET
UDI-Device Identifier04953170141744
UDI-Public04953170141744
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K955404
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOTV-S7H-NA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2022
Initial Report FDA Received Date04/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-