TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 500 MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
|
Back to Search Results |
|
Model Number 500AHCT |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/29/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reported complaint was confirmed.
The srt replaced the arterial bpm probe to resolve the failure.
If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
The service repair technician (srt) reported that during testing of the device at the service center, the arterial blood parameter monitor (bpm) probe failed the high potential (hi-pot) test.
There was no patient involvement.
|
|
Search Alerts/Recalls
|
|
|