• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. 4.5MM INCISOR PLUS PLATINUM BLADE; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. 4.5MM INCISOR PLUS PLATINUM BLADE; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72203013
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2022
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that during meniscectomy, after opened a platinum incisor 4.5mm as per his usual process and started to resect meniscal tissue, it was noticed 'metal shavings' on the meniscal surface.They were shiny and metallic in appearance which were suctioned away.The procedure was completed with non-significant delay using the same faulty device.No other complications were reported.
 
Manufacturer Narrative
H10: h3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The inner surface of the distal end of the outer shaft has small areas of damage from the inner shaft impacting the outer shaft.A functional assessment of the device found the device makes scraping sounds when connected to the reference unit.A complaint history review found no similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include insufficient irrigation or lack of irrigation.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM INCISOR PLUS PLATINUM BLADE
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14166395
MDR Text Key289731850
Report Number1219602-2022-00582
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010655660
UDI-Public03596010655660
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72203013
Device Catalogue Number72203013
Device Lot Number50970235
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-