As reported, approximately one year post op the initial tsa, this male patient's was revised due to humeral liner disassociation.Patient was performing overhead activity 2 weeks ago and felt something in his shoulder.Presented to clinic with pain and x-ray showing possible liner disassociation.He was revised yesterday by surgeon to a new o tray and 46 mm liner of same size.No further information available.Explant is not given to rep without patient permission.Patient last known to be in stable condition.
|
|
Section h10: (d4) catalog number: 320-46-00, serial number: (b)(6), expiration date: 20-aug-2024, unique identifier (udi) #: (b)(4).(h3) the revision reported was likely the result of incomplete seating of the liner during implantation, bone impingement, patient-related conditions, a post traumatic event, or any combination of these possibilities, which led to humeral liner disassociation.However, this cannot be confirmed as the devices were not returned for evaluation.(h4) device manufacture date: 26-aug-2019 section h11: the following sections have corrected information: (d1) brand name: equinoxe reverse 46mm humeral liner +0 (d2b) common device name: prosthesis, shoulder, non-constrained, metal/polymer cemented.
|