| Model Number |
01526 |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A complaint handler followed up with the customer (b)(6) 2022.And she stated that she had been using the symphony pump since (b)(6) a few times a day.At the time of the incident she said she was not in the room where it was plugged in and then her husband called her because he smelled something.She did not see any fire and she indicated that she did not have any issues with her outlet that she is aware of.Additionally she stated that there was no heater by the pump.The device was evaluated on 4/13/2022.It was determined that the housing was melted and charred.It was also noted that all the internal printed circuit boards (pcb's) were functional.A portion of the power cord had melted but was functional.All charring and melting indicates an external heat source but this could not be verified by the customer.The pump was functional and able to be tested and passed the power test and passed the suction and cycle specifications.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022, a medela account called and alleged that a customer brought a symphony breast pump in and said that it sparked and caught on fire.The unit was showing signs of burning on one side and it was replaced for the customer.
|
| |
|
Search Alerts/Recalls
|