LIVANOVA USA INC SMART TUBING AND CONNECTORS; TUBING, PUMP, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number 627368401
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient id, age, gender, weight and ethnicity: no patient information has been provided.Livanova usa inc manufactures the complained circuit.The incident occurred in (b)(6).Follow up with the customer clarified that there is only one unit affected by the mis-assembly.In detail, the red and blue tapes were opposite at the connectors as they were at the soft y, so what the surgeon thought was the red line back at the pump, the perfusionist thought it was the blue line.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event or Problem Description
Livanova usa inc has received a report that, after a few times of trying to give cardioplegia and not being successful and the surgeon noticed that the red and blue tapes were not the same on opposite ends of the tubing.Once discovered they were able to work things out and complete the procedure.There is no report of any patient injury.
 
Event or Problem Description
See intial report.
 
Additional Manufacturer Narrative
The complained device has been received at livanova usa.Initial visual inspection confirmed the reported condition: two of the red and blue tapes were switched on the two legs.Per drawing for 627368401, one leg is supposed to have 2 red tapes on one leg, and 2 blue tapes on the other leg.However, each leg had one red and one blue tape.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event or Problem Description
See intial report.
 
Additional Manufacturer Narrative
Livanova received a report stating that, during procedure, no cardioplegia could be administered due to misassembly.No patient/user affected.The complained unit was returned to livanova for investigation.Visual inspection confirmed the reported condition: two of the red and blue tapes were switched on the two legs.Per relevant technical documentation, one leg is supposed to have 2 red tapes on one leg, and 2 blue tapes on the other leg.However, each leg had one red and one blue tape.The dhr review highlighted that the lot whose claimed product belong to was released as conform according to specifications.By reviewing the complaints database, no further similar event has been recorded on this pts pack code.Based on above, most likely the root cause was due to a manufacturing operator error during the assembly phase of this pts and then he didn't follow tape down.To prevent reoccurrence, the capa-arv-2022-0003 was initiated.The risk is in the acceptable region.No other corrective action is deemed necessary.Livanova will keep monitoring the market.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMART TUBING AND CONNECTORS
Common Device Name
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA USA INC
14401 w 65th way
arvada CO 80004
Manufacturer (Section G)
LIVANOVA USA INC
14401 w 65th way
arvada CO 80004
Manufacturer Contact
enrico greco
14401 w 65th way
arvada, CO 80004
MDR Report Key14166757
Report Number1718850-2022-00035
Device Sequence Number12018153
Product Code DWE
UDI-Device Identifier00803622133671
UDI-Public(01)00803622133671(240)627368401(17)240131(10)2201900053
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K981613
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2024
Device Catalogue Number627368401
Device Lot Number2201900053
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2022
Supplement Date Received by Manufacturer04/28/2022
06/30/2022
Initial Report FDA Received Date04/21/2022
Supplement Report FDA Received Date05/27/2022
07/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-