Event date is unknown date in 2022.Initial reporter is j&j company representative.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that tip for dhs/dcs impactor (338.28) was observed that the handle was broken and the broken part was not returned.There were scratches on the device but have no impact on the device functionality.No other issues were identified with the returned device.A dimensional inspection for the tip for dhs/dcs impactor (338.28) was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces as mentioned in the event description.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the tip for dhs/dcs impactor (338.28) would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: no issues with drawings dimensional inspection: n/a.It was reported that dhs impactor has been damaged throughout years of use.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device from attachment.Visual analysis of the photo revealed that ins f/dhs/dcs impactor no.338.280 was deformed and scratched due to use.The overall complaint was confirmed for ins f/dhs/dcs impactor no.338.280.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history: part# 338.26, synthes lot# h786340, supplier lot# h786340, release to warehouse date: 15 oct2019, supplier: avalign techologies-nemcomed.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|