• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES SWITZERLAND SAR CODMAN VPV PROGRAMMER REVISED; VPV SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES SWITZERLAND SAR CODMAN VPV PROGRAMMER REVISED; VPV SYSTEM Back to Search Results
Model Number 823192R
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A facility reported a codman vpv programmer was overheating when attempting reading for patient¿s hakim valve.No additional information available.
 
Additional Manufacturer Narrative
The programmer was returned for evaluation: dhr - lot met specifications when released.Failure analysis - fault confirmed.The transmitter cover is cracked and must be replaced, along with the attached switch and dome plug.While testing the transmitter, it did eventually overheat.However, this was after ten adjustments, which is a normal number a transmitter operating in cool ambient temperature.Transmitter cover, switch and dome plug have been replaced.Repair performed with reset, safety, and functional test.The possible root cause of the defective device could be due to a bad handling from the user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN VPV PROGRAMMER REVISED
Common Device Name
VPV SYSTEM
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key14167038
Report Number3013886523-2022-00181
Device Sequence Number14708670
Product Code JXG
UDI-Device Identifier10886704078609
UDI-Public10886704078609
Combination Product (Y/N)N
PMA/510(K) Number
K061876
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number823192R
Device Catalogue Number823192R
Device Lot Number133030
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/18/2022
Supplement Date Received by Manufacturer05/31/2022
Initial Report FDA Received Date04/21/2022
Supplement Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/06/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-