Model Number 03.033.001
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/17/2022
Event Type  malfunction  
Manufacturer Narrative
Additional narrative: initial reporter is a j&j employee.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
On (b)(6) 2022, it was reported that when striking the threaded driving cap to insert the frn, the threaded portion of the driving cap broke off into the insertion handle.There was patient involvement.No surgery delayed due to the reported event.There was no surgical delay.The procedure was successfully completed.There was no patient consequence reported.This report is for one (1) radiolucent insertion handle frn.This is report 1 of 2 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIOLUCENT INSERTION HANDLE FRN
Type of Device
ROD, FIXATION, INTRAMEDULLARY
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14167137
MDR Text Key289746939
Report Number2939274-2022-01416
Device Sequence Number1
Product Code HSB
UDI-Device Identifier10886982271051
UDI-Public(01)10886982271051
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172157
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number03.033.001
Device Catalogue Number03.033.001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DRIVING CAP/THREADED
-
-