| Model Number |
101/870/075CZ |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the use of the product, the pilot balloon was found detached from the inflation line.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: h6: event problem and evaluation codes: updated.H10: device evaluation: one used decontaminated sample 10009163-003 7.5mm deht blu suction aid sub-assy was returned for investigation without its original packaging (see photo of sample).Under visual inspection we noticed that inflation line was detached from pilot balloon.10001429-003 infl.Line s/assy soft-seal 7.5mm is assembled in smiths medical tijuana therefore secondary investigation object was created with purpose to capture component investigation.Similar customer complaints have been recently received therefore internal non-conformity report ncr-000928 was created with purpose to evaluate this issue.Smiths medical tijuana site confirmed that no more physical complaint samples are needed for investigation therefore sample is kept in smiths medical czech republic.The cause of the reported problem was traced to the manufacturing process.A dhr review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.
|
| |
|