| Model Number |
N/A |
| Medical Device Problem Codes |
Thermal Decomposition of Device (1071); Corroded (1131); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the connection issue was due to a corroded/burnt nose board.The event occurred during cleaning.There was no patient harm reported.No adverse events were reported as a result of this malfunction.
|
| |
|
Event or Problem Description
|
|
No additional event information available.
|
| |
|
Additional Manufacturer Narrative
|
|
This event has been recorded by zimmer biomet under cmp-(b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined the unit was not connecting to carts.The coupler nose board was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|